Home | Volume 41 | Article number 236

Short communication

Analytical performances of Exacto® HIV self-test in the Central African Republic

Analytical performances of Exacto® HIV self-test in the Central African Republic

Christian Diamant Mossoro-Kpinde1,2,&, Christelle Bobossi1, Serge Tonen-Wolyec3, Ginette Claude Mireille Kalla4, Coretha Baguida-Bokia1, Simplice Sombot-Ndicki1, Gerard Gresenguet2, Francois-Xavier Mbopi-Keou4, Laurent Belec5

 

1Laboratoire National de Biologie Clinique et de Santé Publique, Bangui, République Centrafricaine, 2Faculté des Sciences de la Santé, Université de Bangui, Bangui, République Centrafricaine, 3Faculté de Médecine et de Pharmacie, Université de Kisangani, Kisangani, République Démocratique du Congo, 4Faculty of Medicine and Biomedical Sciences, University of Yaounde I, Yaounde, Cameroon, 5Laboratoire de Virologie, Hôpital Européen Georges Pompidou, Assistance Publique-Hôpitaux de Paris (AP-HP), Université de Paris, Paris Sorbonne Cité, Paris, France

 

 

&Corresponding author
Christian Diamant Mossoro-Kpinde, Laboratoire National de Biologie Clinique et de Santé Publique, Bangui, République Centrafricaine

 

 

Abstract

We herein evaluated the analytical performances of the CE-IVD capillary blood Exacto® HIV self-test (Biosynex, Strasbourg, France) in the Central African Republic (CAR). A cross-sectional study was conducted on a representative national panel of 200 sera positive for HIV and 200 negative for HIV, randomly selected thorough the CAR for HIV seroprevalence surveillance survey, according to reference test. The Exacto® HIV self-test showed 99.5% (95% CI: 98.2-99.9) sensitivity and 100.0% (95% CI: 99.0-100.0) specificity. The Youden´s J index and Cohen´s K coefficient were 0.995. At HIV-1 seroprevalence of 3.5% in the general adult population of the CAR, the positive and negative predictive values were 100% (95% CI: 99.0-100) and 99.9% (95% CI: 98.9-100), respectively. The results are within the limits required by the WHO (i.e. sensitivity ≥ 99.0% and specificity ≥ 98.0%), making Exacto® HIV self-test suitable for routine use in the CAR.

 

 

Introduction    Down

The Central African Republic (CAR) has an adult HIV prevalence rate of 3.5% according UNAIDS´ updated data, with approximately 130,000 people living with HIV-1 [1]. Of these, approximately two-third are unaware of their HIV status [1]. Opportunities for HIV testing could be enhanced by offering accepted and effective strategies such as HIV self-testing in populations that fear stigma and discrimination when accessing conventional HIV counselling and testing in health care facilities [2, 3].

The CAR is continually carrying out new technologies for HIV testing in order to support the national HIV/AIDS policy. A recent study among lay users living in Bangui, the capital city of CAR, previously reported a high rate of the acceptability and the usability of the capillary blood-based Exacto® HIV self-test (Biosynex, Strasbourg, France) [4]. Indeed, the Exacto® HIV self-test is a third-generation immunochromatographic test, using a combination of a specific antibody binding protein that is conjugated to colloidal gold dye particles and synthetic antigens (gp41, gp36) able to detect antibodies (IgG and IgM) against HIV-1 or HIV-2 in whole-blood, serum or plasma, which are bound to the solid-phase membrane [3]. We herein investigated the analytical performances of the Exacto® HIV self-test in the CAR, as recommended by the World Health Organization (WHO) before use [5].

 

 

Methods Up    Down

Study design and setting: this diagnostic performance evaluation was a cross-sectional study including plasma samples collected during the annual HIV seroprevalence surveillance survey from consenting adults from the general population from 2019. Eight field sites in Bangui (Centre National de Références des Maladies Sexuellement Transmissibles et du SIDA, Complexe Pédiatrique, Hôpital de l´Amitié, Hôpital Communautaire, Centre de Santé de Castor, Centre de Santé de Lakouanga, Centre de Santé de Combatant, Centre de Santé de Ngaragba), the capital city of the CAR and seven health care centers (Berberati, Bimbo, Bossangoa, Bozoum, Bria, Damara, Mbaïki) of the main provincial cities of the country were drawn by the Ministry of Health for HIV seroprevalence surveillance survey.

Sampling: according to Centers for Disease Control and Prevention (CDC) and WHO recommendations, the minimum sample size for the study was estimated to be 200 HIV-positive and 200 HIV-negative plasma samples [6]. After listing all plasma samples that were previously positive or negative with the reference test, random sampling was used to select the 200 positive and 200 negative plasma samples to be tested with the Exacto® HIV self-test.

Laboratory procedures: laboratory evaluation of the Exacto® HIV self-test was carried out at the Laboratoire National de Biologie Clinique et de Santé Publique of Bangui. Indeed, a reference panel of plasma samples (200 positive and 200 negative) which had been tested according to reference national serological algorithm for HIV testing, using in parallel Genscreen ULTRA HIV Ag-Ab HIV-1/2 Version 2 (Bio-Rad, Marnes-la-Coquette, France) and Murex HIV 1.2.0 Ag/Ab Combination (Diasorin, Saluggia, Italy) was further tested with the Exacto® HIV self-test. Exacto® HIV self-test was performed using 5 μL of plasma without buffer, following the instructions of the manufacturer. The results of the study test were read by two clinical microbiologists blinded regarding the reference test results. Indeterminate readings were further read by a third microbiologist. Inconclusive plasmas were further tested by real time RT-PCR (Roche Molecular Systems, Inc., Branchburg, NJ, USA) for final diagnosis.

Ethical considerations: the study was approved by the Ethical and Scientific Committee, Faculty of Health Sciences of Bangui, constituting the Institutional Ethical Committee.

Statistical analysis: data were entered into an Excel database (Redmond, WA, USA) and analysed using SPSS 20.0 (Chicago, IL, USA). Analytical performances of Exacto® HIV self-test was defined using sensitivity, specificity, concordance, accuracy, positive predictive values (PPV) and negative predictive values (NPV) as previously described [7]. Briefly, the Cohen´s Kappa coefficient calculation was used to estimate the concordance, and interpreted according to the Landis and Koch scale; accuracy was estimated by Youden´s J index; the PPV and NPV were calculated by using Bayes´ formulae taking into account of the HIV prevalence of 3.5% in the CAR.

 

 

Results Up    Down

The results of the analytical performances of Exacto® HIV self-test are depicted in Table 1. Among the 200 HIV samples known to be HIV-positive with reference test, only 1 was negative by Exacto® HIV self-test, but positive by RT-PCR, and was thus considered as false negative result. Thus, a total of 199 plasma samples of 200 true-positive plasma samples were found positive by Exacto® HIV self-test, while 200 plasmas of the 200 true-negative plasmas samples were all found negative with Exacto® HIV self-test. Taken together, the sensitivity and specificity of the Exacto® HIV self-test were 99.5% (95% CI: 98.2-99.9), and 100% (95%CI: 99.0-100), respectively.

The reliability of Exacto® HIV self-test estimated to 99.5% (95% CI: 98.2-99.9) by the Cohen´s K coefficient measuring the concordance between the HIV self-test and reference serological algorithm, demonstrating almost perfect agreement according to Landis and Koch scale. The accuracy of Exacto® HIV self-test estimated to 99.5% (95% CI: 98.2-99.9) by the Youden's J index, demonstrating excellent accuracy. At HIV-1 seroprevalence of 3.5% in general adult population of the CAR in 2019, the positive and negative predictive values were 100% (95% CI: 99.0-100) and 99.98% (95% CI: 98.9-100), respectively.

 

 

Discussion Up    Down

We herein evaluated the analytical performances of the Exacto® HIV self-test (Biosynex, Strasbourg, France) in the CAR on a representative national panel. The Exacto® HIV self-test showed high sensitivity (99.5%), specificity (100%), concordance, accuracy and positive and negative predictive values.

The virological performances of HIV self-test in French-speaking countries in sub-Saharan Africa are yet poorly established. In practice, clinical validation of the analytical performances of HIV self-test in real-life should be always carried out in a sufficiently large number of target population subjects before introducing them into the routine diagnosis, as strongly recommended by the WHO [8].

In addition, Central Africa is characterized by broad genetic diversity of HIV-1 strains, which can be associated with false negativity of HIV immunochromatographic rapid diagnostic tests [9], and by a variety of factors which can be associated with false positivity or unspecific reactivities, including disturbances affecting the B cell-driven immunity during infectious diseases, such as marked immunological stimulation, strong non-specific polyclonal B-cell activation, hypergammaglobulinemia, and production of circulating immune complexes [10].

HIV self-tests are more and more developed and may be of variable quality [7]. In the present study, the analytical performances of Exacto® HIV self-test in the CAR, were evaluated using a collection of 200 positive and 200 negative sera randomly selected thorough the CAR for HIV seroprevalence surveillance survey. The results showed excellent analytical performances of the Exacto® HIV self-test, despite the risk of false-positive results with frequent inconclusive sera in this area of Africa [5, 10]. Only one positive for HIV was found negative by the HIV self-test, but the final estimated sensitivity remained high. Finally, the analytical performances of Exacto® HIV self-test were within the limits required by the WHO for HIV self-testing i.e. sensitivity ≥ 99.0% and specificity ≥ 98.0% [8], likely allowing it to detect HIV-1 strains circulating in the CAR. A previous study conducted in the Democratic Republic of the Congo showed similar high analytical performances of the Exacto® HIV self-test [3].

This study has some limitations. This study did not consider the possibility of misinterpretation by lay users, as previously described. Indeed, under field conditions, misinterpretation of the HIV self-test could alter the proportion of falsely interpreted positive and negative results, thus reducing its analytical performance in the hands of lay users.

 

 

Conclusion Up    Down

In conclusion, the Exacto® HIV self-test showed high sensitivity and specificity, making the test within the analytical limits required by the WHO to be used in routine in the general population for HIV testing in CAR. Taken together, the high analytical performances in addition to the previously reported high practicability and usability of the Exacto® HIV self-test make the test suitable for routine use in the CAR.

What is known about this topic

  • Field validation of capillary HIV Self-Test constitutes a mandatory prerequisite before use;
  • Broad spectrum of HIV-1 strains are circulating in Central Africa;
  • HIV rapid test may lack specificity in the Central African context.

What this study adds

  • Exacto® HIV Self-Test is highly sensitive and specific in Central Africa;
  • Exacto® HIV Self-Test can be used safely by adult profane public with high confidence in Central Africa.

 

 

Competing interests Up    Down

The authors declare no competing interests.

 

 

Authors' contributions Up    Down

CDMK, STW, GG and LB conceived the study; CDMK, CB, CBB and SSN carried out the experimentations; STW and GCMK made statistical analyses; CDMK, STW, FXMK and LB wrote the paper. All authors have read and approved the final version of the manuscript.

 

 

Acknowledgments Up    Down

The authors are grateful to Biosynex, Strasbourg, France, for providing for the tests for the study. Dr Serge Tonen-Wolyec was recipient of ERASMUS program between the University of Kisangani, Democratic Republic of the Congo, and the University of Liège, Belgium.

 

 

Table Up    Down

Table 1: analytical performances of Exacto® HIV self-test

 

 

References Up    Down

  1. UNAIDS. Country factsheets. Central African Republic. Accessed December 11th 2020.

  2. Longo JDD, Diemer HSC, Tonen-Wolyec S, Mboumba Bouassa RS, Tépka G, Grésenguet G and Bélec L. HIV Self-Testing In Central Africa: Stakes And Challenges. Health Sci Dis. 2018;19(2):1-8. Google Scholar

  3. Tonen-Wolyec S, Batina-Agasa S, Muwonga J, N´kulu FF, Mboumba Boussa RS and Bélec L. Evaluation of the practicability and virological performance of finger-stick whole-blood HIV self-testing in French-speaking sub-Saharan Africa. PLoS One. 2018 Jan 10;13(1):e0189475. PubMed | Google Scholar

  4. Grésenguet G, Longo JDD, Tonen-Wolyec S, Mboumba Bouassa RS and Belec L. Acceptability and Usability Evaluation of Finger-Stick Whole Blood HIV Self-Test as an HIV Screening Tool Adapted to The General Public in The Central African Republic. Open AIDS J. 2017;11:101-118. PubMed | Google Scholar

  5. World Health Organization (WHO). Guidelines on HIV self-testing and partner notification: supplement to consolidated guidelines on HIV testing services. Accessed December 11th 2020.

  6. CDC, WHO and APL. Guidelines for Appropriate Evaluations of HIV Testing Technologies in Africa. Accessed December 10th 2020.

  7. Tonen-Wolyec S, Muwonga Masidi J, Kamanga Lukusa LF, Nsiku Dikumbwa G, Sarassoro A, Bélec L. Analytical performances of Exacto Test HIV self-test: A cross-sectional field study in the Democratic Republic of the Congo. Open Forum Infect Dis. 2020 Nov 29;7(12):ofaa554. PubMed | Google Scholar

  8. World Health Organization (WHO). HIV Self-Testing Strategic Framework: A Guide For Planning, Introducing And Scaling Up HIV Testing Services. Accessed December 11th 2020.

  9. Mossoro-Kpinde CD, Gody JC, Mboumba Bouassa RS, Mbitikon O, Jenabian MA, Robin L et al. High levels of virological failure with major genotypic resistance mutations in HIV-1-infected children after 5 years of care according to WHO-recommended 1st-line and 2nd-line antiretroviral regimens in the Central African Republic. Med (Baltimore). 2017 Mar;96(10):e6282. PubMed | Google Scholar

  10. Klarkowski D, O´Brien DP, Shanks L, Singh KP. Causes of false-positive HIV rapid diagnostic test results. Expert Rev Anti Infect Ther. 2014 Jan;12(1):49-62. PubMed | Google Scholar